Two-Dimensional Separation of Protein Samples
Separation of labeled proteins is carried out using traditional 2-D polyacrylamide gel electrophoresis (chapters 2โ4: EF using Ettanโข IPGphorโข 3 Isoelectric Focusing System and accessories; Electrophoresis conditions with precast gels for both Ettanโข DALTsix and Ettanโข DALTtwelve; Running a MultiPhorโข II protocol).
Protocol
A. Combining protein samples for multiplexing
Protein samples labeled with different CyDyeโข DIGE Fluor minimal dyes are combined according to the experimental design. For best results, one or two labeled protein samples (usually Cy3 or Cy5) are combined with a labeled internal standard (usually Cy2), which is a pool of aliquots of all biological samples in the experiment.
B. Diluting labeled protein sample in sample buffer
The sample mixture is diluted further in sample buffer prior to separating the individual proteins on a 2-D gel.
Ettanโข DIGE system-compatible 2ร sample buffer contains 7 M urea, 2 M thiourea, 2% CHAPS (w/v), 2% IPG buffer or Pharmalyteยฎ (v/v) for IEF, 2% DTT (w/v).
Add 1 volume of 2ร sample buffer to sample. Mix and leave on ice for at least 10 min.
C. Rehydrating Immobilineยฎ DryStrip gel
Refer to sections 2.4โ2.7 for a discussion of rehydration and sample application methods.
Ettanโข DIGE system-compatible rehydration buffer contains 7 M urea, 2 M thiourea, 4% CHAPS (w/v), 1% IPG buffer or Pharmalyteยฎ (v/v) for IEF, 2% (w/v) DTT.
D. Separating proteins in the first dimension
Ettanโข IPGphorโข 3 Isoelectric Focusing System and MultiPhorโข II Electrophoresis System are both suitable for Ettanโข DIGE system applications. Detailed instructions for use of the systems are given in chapters 2 (IEF using Ettanโข IPGphorโข 3 Isoelectric Focusing System and accessories) and 4 (First-dimension IEF using MultiPhorโข II Electrophoresis System and Immobilineยฎ DryStrip Kit), respectively. Ettanโข DIGE applications are described in detail in the Ettanโข DIGE user manual.
E. Separating proteins in the second dimension
Low-fluorescence glass plates must be used for gels used in Ettanโข DIGE system. Standard glass or plastic-backed plates can result in the generation of a high background signal.
DIGE Gel is a 12.5% precast, low-fluorescent polyacrylamide gel cast in a low-fluorescent glass cassette specially developed for 2-D DIGE analysis. DIGE Gel should be used with the DIGE Buffer Kit, which consists of concentrated running buffers and Sealing Solution for attaching Immobilineยฎ DryStrip Gels (IPG strips) to the top of the polyacrylamide gel. The capacity of the buffer system is similar to the commonly used Laemmli (Tris-Glycine) buffer system, and the separation performance of DIGE Gel is comparable to 12.5% Laemmli gels.
DALT gels are large enough to accommodate the longest Immobilineยฎ DryStrip gels (24 cm) and can be run in batches of up to 12 gels at a time. DALT gels can also be poured using the DALT gel caster. DALT gels can also be run using the DIGE buffer kit. Detailed protocols for gel casting can be found in the Ettanโข DIGE user manual, Ettanโข DALTtwelve user manual, and Ettanโข DALTsix user manual.
The procedures for equilibrating strips, positioning them, and electrophoresis are identical to those for standard 2-D analysis. Refer to sections 3.1 (Equilibrating Immobilineยฎ DryStrip gels) and 3.3 for details.
If the gels are to be scanned immediately, store the gels in SDS electrophoresis running buffer at room temperature in a light-tight container. Scan the gels as soon as possible as the protein spots in the gel will diffuse with time. If the gels cannot be scanned on the day of running, they should be stored in the dark at +4 ยฐC and kept moist. Remember to let the gels warm to room temperature before scanning because temperature affects the fluorescent signal.
Do not fix the gels prior to scanning as this will affect quantitation of the labeled protein spots.