HPLC Analysis of 16 Triazine and Phenyl Urea Compounds on Ascentisยฎ Express RP-Amide

HPLC Analysis of 16 Triazine and Phenyl Urea Compounds on Ascentis® Express RP-Amide application for HPLC

Materials

๊ด€๋ จ ์ œํ’ˆ

์ œํ’ˆ ๋ฒˆํ˜ธ
์„ค๋ช…
๊ฐ€๊ฒฉ

Atrazine

PESTANALยฎ, analytical standard

Terbumeton-desethyl

PESTANALยฎ, analytical standard

Atrazine-desisopropyl

PESTANALยฎ, analytical standard

Isoproturon

PESTANALยฎ, analytical standard

Linuron

PESTANALยฎ, analytical standard

Diuron

PESTANALยฎ, analytical standard

Chlortoluron

PESTANALยฎ, analytical standard

Cyanazine

PESTANALยฎ, analytical standard

Terbumeton

PESTANALยฎ, analytical standard

Terbuthylazin-desethyl

PESTANALยฎ, analytical standard

Ascentisยฎ Express RP-Amide, 2.7 ฮผm HPLC Column

2.7 μm particle size, L ร— I.D. 15 cm ร— 3 mm

Carbendazim

analytical standard

CONDITIONS

column

Ascentis Express RP-Amide, 15 cm x 3 mm I.D., 2.7 μm particles (53919-U)

mobile phase

[A] water; [B] acetonitrile

gradient

5 to 53% B in 70 min

flow rate

0.4 mL/min

column temp.

35 ยฐC

detector

UV, 240 nm

injection

5 μL

sample

20 mg/L (except terbuthylazin desethyl and terbuthylazine 60 mg/L) in 60:40, water: acetonitrile

์„ค๋ช…

๋ถ„์„ ๋ฉ”๋ชจ

This application demonstrates the suitability of Ascentis Express RP-Amide for the efficient separation of 16 triazine and phenyl urea compounds. Lower column temperature can improve resolution of the critical pair.