콘텐츠로 건너뛰기
Merck

GC Analysis of Trichloroanisole (TCA) in Wine on SLB®-5ms after SPME using 100 μm PDMS Fiber

GC Analysis of Trichloroanisole (TCA) in Wine on SLB®-5ms after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

Materials

used together

제품 번호
설명
가격

SPME fiber assembly Polydimethylsiloxane (PDMS)

metal alloy fiber, df 100 μm(PDMS, needle size 23 ga, for use with autosampler

관련 제품

제품 번호
설명
가격

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

CONDITIONS

sample/matrix

red wine spiked with 5 ppt TCA

SPME fiber

100 μm PDMS metal fiber (57928-U)

extraction

headspace, room temp (30 min)

condition

3 min. at 250 °C

desorption process

250 °C for 3 min

column

SLB®-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)

oven

60 °C (2 min), 8 °C/min to 200 °C (10 min)

MSD interface

300 °C

scan range

SIM, m/z = 195, 197, 210, 212

carrier gas

helium, 1 mL/min. constant

liner

0.75 mm I.D. SPME

설명

법적 정보