GC Analysis of Pesticides in Beef Kidney on SLBยฎ-5ms (20 m x 0.18 mm I.D., 0.36 ฮผm) after QuEChERS Cleanup using Supelโข Que Z-Sep+, MS in SIM Mode, Fast GC Analysis

CONDITIONS
sample preparation
Dispersive (QuEChERS)
sample/matrix
10 gm homogenized beef kidney spiked at 50 ng/g with pesticides
extraction tube
Supel Que Acetate (AC) Tube (55234-U)
extraction process
add 10 mL of acetonitrile; shake for 1 minute; add contents of Supel Que acetate extraction tube (55234-U); shake immediately for 1 minute; centrifuge at 3200 rpm for 5 minutes; (Sample Pretreatment: Homogenize 10 gm beef kidney and spike with pesticides at 50 ng/g.)
clean-up tube
Supel QuE Z-Sep+ tube (55296-U)
clean-up process
transfer 3 mL of the acetonitrile layer into a Supel QuE Z-Sep+ cleanup tube (55296-U); shake for 1 minute; centrifuge at 3400 rpm for 3 minutes; draw off supernatant for LVI-GCMS analysis
column
F.S. CAP. SLB-5MS 20 m x 0.18 mm I.D., 0.36 μm (28576-U)
oven
70 ยฐC ( 0.5 min), 25 ยฐC/min to 125 ยฐC, 10 ยฐC/min to 200 ยฐC, 5 ยฐC/min to 300 ยฐC/min (1 min)
inj. temp.
programmed, 60 ยฐC (0.28 min), 600 ยฐC/min to 325 ยฐC (5 min)
detector
MSD
MSD interface
325 ยฐC
carrier gas
helium, 1 mL/min constant flow
injection
10 μL LVI, PTV solvent vent, rapid injection speed
liner
4 mm ID FocusLiner with taper
sample
Beef Kidney extract
์ค๋ช
๋ถ์ ๋ฉ๋ชจ
Beef kidney was analyzed for pesticides by GC/MS-SIM after extraction and cleanup. Dispersive SPE was used for both extraction and cleanup. Since beef kideny contains approximately 5% fat, Z-Sep + was used for cleanup.
๋ฒ์ ์ ๋ณด
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC