콘텐츠로 건너뛰기
Merck

GC Analysis of Organophosphorus Pesticides (OP-Pest) on a 5% Phenyl Column after SPME using 100 μm PDMS Fiber

GC Analysis of Organophosphorus Pesticides (OP-Pest) on a 5% Phenyl Column after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

Materials

used together

제품 번호
설명
가격

SPME Fiber Assembly

100 μm PDMS, Fused Silica (1 cm), needle size 24 ga, Manual Holder, pk of 3, red hub

관련 제품

제품 번호
설명
가격

SLB®-5ms Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

CONDITIONS

sample/matrix

50 ppb each in 1.5 mL saturated salt water, pH 7.2

SPME fiber

polydimethylsiloxane, 100 μm (57300-U)

extraction

20 min immersion with rapid stirring

column

5% phenyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)

oven

60 °C (1 min) to 300 °C at 12 °C/min, hold 5 min

inj. temp.

splitless (0.75 mm I.D. liner), 270 °C

detector

MS (m/z = 45-400, 0.6sec/scan)

carrier gas

helium, 30 cm/sec