GC Analysis of Aromatics in Reformulated Gasoline on SUPELCOWAX 10

GC Analysis of Aromatics in Reformulated Gasoline on SUPELCOWAX 10 suitable for GC

Materials

๊ด€๋ จ ์ œํ’ˆ

์ œํ’ˆ ๋ฒˆํ˜ธ
์„ค๋ช…
๊ฐ€๊ฒฉ

SUPELCOWAX 10 Capillary GC Column

L ร— I.D. 60 m ร— 0.25 mm, df 0.25 ฮผm

CONDITIONS

column

SUPELCOWAX 10, 60 m x 0.25 mm I.D., 0.25 μm (24081)

oven

50 ยฐC (3 min), 15 ยฐC/min to 265 ยฐC (5 min)

inj. temp.

250 ยฐC

detector

FID, 265 ยฐC

carrier gas

helium, 30 cm/sec

injection

0.5 μL, 200:1 split

liner

4 mm I.D., splitless type, single taper wool packed FocusLinerโ„ข design

sample

reformulated unleaded gasoline

์„ค๋ช…

๋ถ„์„ ๋ฉ”๋ชจ

This application illustrates the analysis of a sample of reformulated gasoline on the SUPELCOWAX 10 column. Due to the polar nature of the phase, the aliphatics elute earlier than the aromatics. There is some overlap in the elution ranges, however it is not as severe as that exhibited by a nonpolar column.