GC Analysis of Aromatics in Reformulated Gasoline on SUPELCOWAX 10

Materials
๊ด๋ จ ์ ํ
์ ํ ๋ฒํธ
์ค๋ช
๊ฐ๊ฒฉ
CONDITIONS
column
SUPELCOWAX 10, 60 m x 0.25 mm I.D., 0.25 μm (24081)
oven
50 ยฐC (3 min), 15 ยฐC/min to 265 ยฐC (5 min)
inj. temp.
250 ยฐC
detector
FID, 265 ยฐC
carrier gas
helium, 30 cm/sec
injection
0.5 μL, 200:1 split
liner
4 mm I.D., splitless type, single taper wool packed FocusLinerโข design
sample
reformulated unleaded gasoline
์ค๋ช
๋ถ์ ๋ฉ๋ชจ
This application illustrates the analysis of a sample of reformulated gasoline on the SUPELCOWAX 10 column. Due to the polar nature of the phase, the aliphatics elute earlier than the aromatics. There is some overlap in the elution ranges, however it is not as severe as that exhibited by a nonpolar column.
๋ฒ์ ์ ๋ณด
FocusLiner is a trademark of SGE Analytical Science Pty Ltd.